Brief introduction:
Air temperature or skin temperature of servo-controlled by micro-computer with oxygen concentration monitoring system and humidity control system (optional); integrated sensor box including the skin temperature sensor, air temperature sensor, oxygen concentration sensor and humidity sensor; double wall hood an reduce the heating losing and the effect of outer heat resource; automatic curtain can reduce the heat losing ; inclination of infant bed can be adjusted; quite operation windows can reduce the disturbance to the baby during nursing; touch switch panel with LED digital display, set-temp, real-temp and auxiliary temperature are displayed separately; >37℃ temperature setting; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100%; whole aluminum water tank, and it can be filled with water or discharge; RS 232 connector, data output.
Standard configure:
Hood, infant bed, temperature controller, integrated sensor box, water tank, air filter, Guardrail, I.V. Pole, cabinet.
Technical specification:
Power supply : AC220V,50HZ
Power input: 800VA
Air Temperature control range:25℃~37℃ can override to 39℃
Skin Temperature control range:34℃~37℃ can override to 38℃
Sensor precision: ≤0.3℃
Changing of temperature: ≤0.5℃
Uniformity of temperature: ≤0.8℃
Temperature rising time: ≤30min
Noise level: ≤55dB(A)
Inclination of bassinet: 0°~5°
Indication range of oxygen concentration:0%~99%
setting range of oxygen concentration:20%~60%
Indication precision of oxygen concentration:±3%
Humidity indication range:0%RH~90%RH
Humidity control range:20%RH~90%RH
Humidity indication precision:±6%RH
Alarming:
audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater (except for oxygen concentration alarm, humidity alarm, the low deviation alarm), meanwhile, the auxiliary temperature indication window can indicate the alarm code. For the following alarm, the relevant light is on, and for the alarm 1-4, press the alarm key to cancel the alarming sound for 5min.
Over-temp alarm: when the air temperature control ≤38℃(setting temperature is less than 37℃ ), or≤40℃(setting temperature is more than 37℃);skin temperature control ≤40℃;
Upper Deviation alarm: air temperature control +3.0℃, skin temperature control +1.0℃;
Lower Deviation alarm: air temperature control -3.0℃, skin temperature control -1.0℃;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put outside of incubator;
Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.
oxygen concentration alarm:±5%O2;
humidity deviation alarm:±15%RH。
Package:
Gross Weight:70Kg
Net Weight: 59Kg
Measurement:1085mm×640mm×915mm
Notice:
Some technical data of this model can be changed properly in standard permission range according to the standard and user’s requirement like oxygen concentration monitoring system and humidity control system, lift stand, blue light radiant box, infant scale, tray. If the technical specification of above product changes, please refer to the standard at selling time.