Features:
- 5.5inch color LCD for clear information.
- Optional telemetry transmitter enables both ECG and Respiration monitoring
- Voice Guidance provides the instrument status
- Built-in Resistance indicator on Paddle for efficient energy discharge
- Synchronous/Asynchronous mode selectable
- Rechargeable battery or AC power operation.
- Three steps for simple and quick Operation:
- Select energy level
- Energy Charge
- Discharge
Specifications:
Defibrillator:
Output energy levels(with 50 Ohms load)
2,3,5,7,10,20,30,50,70,100,150,200,300,360joules(with external paddles)
2,3,5,7,10,20,30,50joules(with internal paddles)
Charge times: within 10seconds to 360joules on AC operation
within 12seconds to 360joules on DC operation
Discharge modes: Synchronous and asynchronous
Discharge: Single-phase pulse(with 50 Ohms Load)
Peak Voltage: 5KV max( with 100 Ohms load)
Number of discharge: 30times min(battery operation)
ECG
Leads: I, II, III(ECG Lead), Paddle Lead, telemetry lead
Time Constant: 0.3 seconds
Sensitivities: 1/4, 1/2, 1, 2, 4cm/m V manual or automatic selection
Monitor
Display: 5.5inch LCD
Display mode: Non-fade, Stationary
Cascaded ECG or 2-channel (ECG and Respiration with LX-5120)
Sweep Speed: 25mm/sec.(ECG), 8mm/sec., or 25mm/sec(Respiration)
Display time: Approx. 8.8seconds
Alarm Setting limits: 15 to 300 bpm(HR)
Alarm indication: Audible and visual
Recorder
Recording method: Thermal dot array method
Paper speed: 25mm/sec.
Recording paper: 50mm x 30m
Print Items: Year/Month/Day, Hour:Minute, ECG leads,set energy level,output energy level,sensitivity,Heart rate,synchronous point,discharge point.
Recording Modes: Manual(approximately 8.8seconds delayed waveform),automatic and freeze waveform recording.
Telemetry
Transmitter: Optional LX-5120
Receiver system: Double superheterodyne with crystal controlled synthesizer.
Antenna: Built in (whip antenna)
Reveivable parameters: ECG,RESPIRATION
General:
Protection against electrical shock: Class I, Typ CF (With ECG lead,Internal paddles)
Type BF (With external paddles)
Battery charge time: Approx 150minutes
Battery Operation time: 1hour when operated for monitoring purpose only.
Power requirements: AC 90V to 240V, 5/60Hz, 170VA Max, or 12V DC rechargeable battery.
Dimesions: 33cm x 43cm x 12cm
Weight: Approx. 8.2kgs (incl. external paddles TE-300)