Specifications:
1. General Specifications:
Size: 205mm×200mm×95mm
Weight: 1 kg
Display Specification
Parameter Display: 7-segment LEDs
Information Display: Fixed characters and icons lighted by single-color LEDs
Power Supply Indicator: Dual-color LED
2. Power Supply
AC Power Adaptor
Input: 100-240VAC, 50/60Hz
Output: DC 9V,800mA
Battery
Lithium-ion rechargeable 5AHr 7.2V battery
Operating time: 10 hours with full battery capacity.
(Condition: 25℃, NIBP working period is 10 minutes.)
Recharge Time: 20 hours.
3. Environment
Temperature
Working: 0 ~ 45 (°C)
Transport and Storage: -25 ~ 85 (°C)
Humidity
Working: 30% - 85 %
Transport and Storage: 30% - 95 % (no condensation)
4. Monitoring Parameter
NIBP
Method: Oscillometric
Mode: Manual, Auto, STAT
Measuring Interval in AUTO Mode: 1, 2, 3, 5, 10, 15, 20, 30, 60 (min)
Measuring Period in STAT Mode: 5 min
Cuff Pressure Range: 0 - 300 (mmHg)
Pulse Rate Range: 30 - 250 (bpm)
Measurement Rage: Adult 30-255mmHg, Pediactri 30-160mmHg
Over-pressure Protection: Adult 300mmHg, Pediactri 220mmHg
Pressure Resolution: 1mmHg
The SYSTOLIC and DIASTOLIC values measured using this device shall be equal to those measured by trained medical personnel using cuff and stethoscope. The MEAN value measured using this device is equal to that measured using end artery blood pressure measurement device.
SpO2
SpO2%
Measuring Range: 35 - 99%
Resolution: 1 %
Accuracy: ±2 %(70% - 99%), unspecified (0% - 69%)
Pulse Rate
Measuring and Alarm Range: 0~250bpm
Resolution: 1bpm
Accuracy: ±1bpm
Storage and Review
Storage Method: Store NIBP and SPO2 data each time after one NIBP measurement. Maximum Store Data of 800 groups.
5. Printer (optional)
Type: Thermal microprinter
Record Width: 57 mm
Recording Types: Real-time printing and review data printing
AC Power Adaptor
Input: 100-240V,50/60Hz
Output: DC 5V, 3A