Multi-Functional Anesthesia Machine (model MHJ-IIC)

  • Model NO.:MHJ-IIC

  • Kind:Anaesthesia

  • Group:Adult

  • Source:Oxygen

  • Type of Operating Light:Anaesthesia

  • Adjustment Range of Flow N2o:0.05-10L/Min

  • O2 Flush:35-75L/Min

  • Vaporizer:0.2-5vol%,with Pressure,Flow and Temperature Comp

  • Tidal Volume:100-1200ml

  • Respiratory Monitor: Tidal Volume:0-1500ml

  • Specification:CE ISO

  • HS Code:90189070

  • Type:Anaesthesia

  • Certification:CE, ISO13485

  • LCD Display:With LCD Display

  • Anesthesia:Semi Open Suction

  • Adjustment Range of Flow :O2:0.1-15L/Min

  • Adjustment Range of Flow Orc:O2 Concentration≥25%(O2+N2o)

  • Apl Adjustment:0 -7kpa;CO2 Absorber:1.6L

  • Anesthesia Ventilator: Frequency:6-40bpm

  • Maximum Airway Pressure:6kpa

  • Trademark:everich

  • Origin:China

Product Description

MHJ-IIC is a multi-functional anesthesia machine incorporating anesthesia, respiratory and monitoring functions.
Anesthesia Machine: Two-stage pressure relief; high-precision N2Oand O2 flowmeters.ORC,O2-deficiency alarm and N2O cut-off device. It′s equipped with ZFG-B1 compensation vaporizer, such as Drager 193.,Ohmeda TEC-5 and Penlin Sigma Delta.
Integrated breathing system satisfies clinical demand of low-flow anesthesia.
Anesthesia Ventilator: Pneumatic-driven and electric-controlled, time-cycled and volume-constant.
Low air-consumption. Power failure alarm and low-flow alarm.
Monitor: Equipped with respiratory monitory for realtime monitor on tidal volume, minute volume, respiratory frequency and oxygen concentration assuring patients safety

Specification

Adjustment range of flowO2:0.1-15L/min
 N2O: 0.05-10L/min
ORC: O2 concentration≥25%(O2+N2O)
O2 Flush: 35-75L/min 
Anesthesia ventilator Frequency6-40bpm
Tidal volume100-1200ml
Maximum airway pressure6kpa
Respiratory monitor Tidal volume0-1500ml
MV0-40L/min
 Oxygen concentration0-40 L/min


Adjustment range of flow :O2:0.1-15L/min
                                                    N2O: 0.05-10L/min
                                                    ORC: O2 concentration≥25%(O2+N2O)
O2 Flush: 35-75L/min 
O2 deficiency alarm/N2Ocut-off:
     As O2 pressure is≤0.2Mpa,audible;
     As O2 pressure is≤0.14Mpa,N2O decrease;
     As O2 pressure is≤0.04Mpa,N2O is cut off automatically;
APL adjustment:               0 -7kPa;CO2 absorber:1.6L
Vaporizer:                           0.2-5VOL%,with pressure,flow and temperature compensation functions.
Anesthesia ventilator:
      Frequency:6-40bpm:
     Tidal volume:100-1200ml
     1:E:1:-1:3,adjustable
     Maximum airway pressure:6kpa
Respiratory monitor:
     Tidal volume:0-1500ml
     MV: 0-40L/min
     Oxygen concentration:0-40 L/min
     Airway Pressure:0-9.9kpa