Laptop Multi-Parameter Patient Monitor (PM-900A)

  • Model NO.:PM-900A

  • Type:Pulse Oximeter

  • Group:All People

  • Gross Weight:1kg

  • Power Voltage:AC 100 V-240V 50-60Hz

  • Temperature Operation:0 ~ 50 °c

  • Trademark:everich

  • Origin:China

  • Classification:Imaging Diagnostic Equipment

  • Certification:CE

  • Net Weight:600kg

  • Package Size:150*114*45mm

  • Power Iutput:≤70va

  • Power Condition (DC):Built in Chargeable Ni/Mh Batter

  • Specification:CE, ISO

  • HS Code:90181930

Product Description

Features:  
      7 inch high resolution (800x600) colorful TFT LCD
           Simultaneous display of 4 channel waveform on one screen
           Dual NIBP with over pressure protection
           Analysis of real time ST segment and arrhythmia
           Standard configuration: ECG, NIBP, RESP, SPO2, 1-TEMP, PR, drug dose calculation, OXYCRG
           Optional configuration: Built in thermal printer, 2-TEMP
           Standard RJ-45 socket to support networking capacity of maximum 66 beds and compatibility with HIS
           Digital SPO2 enables effective examination in case of active patient and low infusion as well
                interference caused by electrosurgical unit
           Compact and portable design allows uninterrupted operation with its built in rechargeable battery
Specifications:


 ECG    


Input:3lead ECG cable, standard AAML cable connector
Lead selection: I, II, III
Gain selection: ×1, ×2, ×4, ×8
Sweep speed: 12.5mm/s, 25mm/s,50mm/s
Bandwidth: 0.5Hz~40Hz, (for monitoringor operation)
0.05Hz~100Hz (for diagnosis)
Calibration signal: 1m V
Protection against interference arisen from electrosurgical uni, pace maker and/or defibrillator    


 SPO2    


Range: 5%~100%
Accuracy:±2%(70%~100%), ±3% (40%~69%)

Pulse Rate
Range: 0bpm~250bpm
Accuracy: ±1bpm    


NIBP    


Measurement method: Oscillometric
Measurement mode: Manual, Auto, STAT
Measurement interval in AUTO mode: 1,.3,5,10,15,20,30,60 (min)
Measurement type: SYS, DIA, MEAN
Resolution: 1mmHg
Over pressure protection: Available    


 Temperature    


Measuring range: OºC~50ºC
Resolution:0.1ºC
Accuracy: ±0.1ºC    


Heart Rate    


Measure range: 20bpm~250bpm
Accuracy: ±1bpm
Resolution: 1bpm
Alarm mode: Audible and visible alarm    


Operation Environment    


Temperature: OºC~45ºC
Humidity:30ºC~85ºC(non-condensed)    


Power Condition    


DC: Built in chargeable NI/MH batter
AC power: 100V~24V,50Hz~60Hz,10VA    


 Dimension    


150mm×114mm×45mm    


 Net weight    


600g