12.1 Inch LCD Portable Patient Monitor

  • Model NO.:PM-100A

  • Type:Pulse Oximeter

  • Group:All People

  • Gross Weight:4.5kg

  • Power Voltage:AC 100 V-240V 50-60Hz

  • Temperature Operation:0 ~ 40 °c

  • Resolution:1bpm

  • Specification:CE, ISO

  • HS Code:90181930

  • Classification:Imaging Diagnostic Equipment

  • Certification:CE

  • Net Weight:3.5kg

  • Package Size:318*264*152mm

  • Power Iutput:≤70va

  • Measurement Range:10-300bpm

  • Trademark:everich

  • Origin:China

Product Description

 Features:  
   12.1 inch high brightness colorful TFT LCD
           Simultaneous display of 7 channel waveform by one screen
           Dual NIBP with over pressure protection
           Analysis of real time ST segment and arrhythmia
           Standard configuration: ECG, NIBP, RESP, SPO2, 1-TEMP, PR, drug dose calculation, OXYCRG
           Optional configuration: 12 lead ECG, 2-TEMP, IBP(single/dual channel), ETCO2(side stream/main stgream),
                built in thermal printer and CO
           Standard RJ-45 socket to support networking capacity of maximum 66 beds and compatibility with HIS
           Digital SPO2 enables effective examination in case of active patient and low infusion as well
                interference caused by electrosurgical unit
           Compact and portable design allows uninterrupted operation with its built in rechargeable battery
Specifications:

Size and Weight

Name of the monitorMulti-parameter Monitor
Size318mm×264mm×152mm
Weight4.5Kg
System output

Display type12.1″LCD color display screen, with a resolution of 800×600
Interface with central unitNetwork port
Power supply

Power VoltageAC 100V~240V 50/60Hz
Power Input≤70VA
Fuse1.6A
Safety LevelClass I , Type CF
Battery

TypeRechargeable Lead acid cell, 12V/3.2AH
Charge time≤6 hours
Working hours when charged fully≥30min when using normally
Environment

Ambient TemperatureWorking 0~+40ºC

                       Transportation and storage -20~+50ºC
Relative humidityWorking ≤85%

                       Transportation and storage ≤93%
Atmospheric pressureWorking 860~1060 hPa

                       Transportation and storage 500~1060 hPa
ECG

Lead Mode5-leads ECG input

                       3-leads ECG input

                       10-leads ECG input
Lead selectionI, II, III, aVR, aVL, aVF, V-

                       I, II, III

                       I, II, III, aVR, aVL, aVF, V1~V6
Gain2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Input impedance≥5.0 Mohm (d.c. ~ 60 Hz)
CMRRDIAG ≥85dB

                       MON ≥105dB

                       OPS ≥105dB
Frequency responseDIAG 0.05~75Hz

                       MON 0.5~25Hz

                       OPS 1~15Hz
Electrode offset potential±300 mV d.c. → ±600mV d.c.
Leakage Current<10 uA
ECG signal range±6.0 mV
Baseline recovery<5 s After Defi. (MON or OPS mode)
InsulationBreakdown Voltage 4000VAC 50Hz/60Hz
Indication of electrode separationAvailable
Sweep speed12.5mm/s, 25mm/s, 50mm/s
HR Measuring and Alarm

Range10~300 bpm
Refreshing timePer pulse
Resolution1 bpm
Accuracy±1% or ±1 bpm, whichever is greater
Alarm range10~300 bpm, continuously adjustable between high limit and low limit
Alarm indicationSound and light alarming
NIBP

Way of measurement

                       Automatic oscillometry


                       Range of measurement


                       Adult
SYS30~275 mmHg
DIA10~255 mmHg
MEAN20~255 mmHg


                       Child
SYS30~235 mmHg
DIA10~160 mmHg
MEAN20~170 mmHg


                       Neonate
SYS30~145 mmHg
DIA10~120 mmHg
MEAN20~140 mmHg
Cuff pressure range

                       0~280 mmHg
Resolution

                       1 mmHg
Pressure Accuracy

                       Static

                       Clinical


                       ±2% or ±3 mmHg, whichever is greater

                       ±5 mmHg average error

                       8 mmHg standard deviation
Unit

                       mmHg, kPa
Pulse rate measuring range

                       40 ~ 240 bpm
Inflation time for cuff

                       Less than 40 sec. (standard adult cuff)
Total cycle time

                       20 to 45 seconds typical (dependent on heart rate and motion artifact)
Intervals for cycle measurement time

                       1,2,3,4,5,10,15,20,30,45,60,90 minutes

                       2,4,8 hours
Overpressure Protection

                       Adult

                       Child

                       Neonatal


                       Hardware and software double protections

                       315±10 mmHg

                       255±10 mmHg

                       170±10 mmHg


                       Range of alarm
SYS

                       0~300 mmHg, continuously adjustable between high limit and low limit
DIA

                       0~300 mmHg, continuously adjustable between high limit and low limit
MEAN

                       0~300 mmHg, continuously adjustable between high limit and low limit
Alarm indication

                       Sound and light alarming
Ways of measurement

                       Single, Cycle and Quick (Q.sys)
SpO2

Range of measurement0~100%
Resolution1%
PrecisionAt 70~100%, ± 2%

                       At 35~69%, ± 3%

                       At 0~34%, unspecified
Range of alarm0~100%, continuously adjustable between high limit and low limit
Alarm indicationSound and light alarming
Type of sensorDual bandwidth light emitting diode
PR:

                       Measuring and alarm Range

                       Resolution

                       Accuracy
25~250 bpm

                       1 bpm

                       ±2% or ±1 bpm, whichever is greater
IBP

Range of measurement

                       -45 ~ +300 mmHg
Resolution

                       1 mmHg
Precision

                       ± 4mmHg or 4% of the reading (select the bigger one of the two)
Sensitivity of transducer

                       5uV/V/mmHg, 2%
Impedance of transducer

                       300~3000 Ohm
Bandwidth

                       d.c. ~ 15Hz
Transducer sites

                       Arterial Pressure (ART)

                       Pulmonary Artery Pressure (PA)

                       Left Atrium Pressure (LAP)

                       Right Atrium Pressure (RAP)

                       Central Venous Pressure (CVP)

                       Intracranial Pressure (ICP)


                       Selection of measurement range


                       ART
0~50mmHg

                       50~150 mmHg

                       100~240 mmHg

                       0~300 mmHg

                       AUTO


                       PA

                       CVP

                       LAP

                       RAP

                       ICP
0~20 mmHg

                       0~30 mmHg

                       0~50 mmHg

                       0~80 mmHg

                       AUTO


                       (Among them, the AUTO switches automatically at an interval of 10 mmHg so as to ensure the waveform is at the state most suitable for observation)


                       Range of alarm
SYS

                       -10~300 mmHg, continuously adjustable between high limit and low limit
DIA

                       -10~300 mmHg, continuously adjustable between high limit and low limit
MEAN

                       -10~300 mmHg, continuously adjustable between high limit and low limit
Alarm indication

                       Sound and light alarm
Unit

                       mmHg, kPa
Insulation resistance

                       >5000MΩ/1000V
Breakdown voltage

                       >4000V AC 50Hz/60Hz
TEMP

Range of measurement0.0~50.0ºC
Resolution of display0.1ºC
UnitCelsius (ºC), Fahrenheit (ºF)
Precision±0.1ºC
Connecting cableCompatible with YSI-4400
Range of alarm0.0~50.0ºC, continuously adjustable between high limit and low limit
Alarm indicationSound and light alarming
RESP

MethodImpedance between RA-LL (R-F)
Measuring impedance range0.5 ~3 Ohms
Base line impedance range500~4000 Ohms (50~120 kHz exciting frequency)
Range of measurement0~120 bpm
Precision±2 bpm
Gainx1,x2,x4
Range of alarm0~120 bpm, continuously adjustable between high limit and low limit
Sweep speed6.25mm/s, 12.5mm/s, 25mm/s
Asphyxia1~60s, off
Alarm indicationSound and light indication
Recorder

Speed of paper feeding12.5mm/s, 25mm/s, 50 mm/s (Only valid whit B Type recorder)
Recording medium50mm thermal paper
Recording trackSingle track or tri-track