Baby Incubator in Hospital (model YP-910)

  • Model NO.:YP-910

  • Function:Analgesia

  • Certification:ISO13485

  • Group:Infant

  • Power Input:600va

  • Skin Temperature Control Range:34oc~37oc

  • Changing of Temperature:<=0.5oc

  • Gross Weight:66kg

  • Measurement:1085X640X915mm

  • Specification:CE, ISO, SGS

  • HS Code:90189090

  • Type:Incubator

  • Theory:Electrotherapeutics

  • LCD Display:Without LCD Display

  • Power Supply:AC220V,50Hz

  • Air Temperature Control Range:25oc~37oc

  • Sensor Precision:<=0.3oc

  • Noise Level:<=55db(a)

  • Net Weight:55kg

  • Trademark:everich

  • Origin:Jiangsu

Product Description

Brief introduction:
servo-controlled by micro-computer; quite operation windows can reduce the disturbance to the baby during nursing; inclination of the bassinet can be adjusted, touch switch panel with LED digital display, set-temp, real-temp and auxiliary temperature are displayed separately; multiply self-checking alarms with audible and visible; heating power rate indicates 0-100%; whole aluminum water tank, and it can be filled with water or discharge; RS 232 connector, data output.
Standard configure:
Hood, infant bed, temperature controller, skin temperature sensor, water tank, air filter, Guardrail, I.V. Pole, cabinet.

Technical specification:

Power supply AC220V,50HZ
Power input600VA
Air Temperature control range25ºC~37ºC
Skin Temperature control range34ºC~37ºC
Sensor precision≤0.3ºC
Changing of temperature ≤0.5ºC
Uniformity of temperature≤0.8ºC
Temperature rising time≤30min
Noise level≤55dB(A)
Inclination of bassinet0°~5°
Gross Weight 66Kg
Net Weight55Kg
Measurement1085mm×640mm×915mm


Power supply : AC220V,50HZ
Power input: 600VA
Air Temperature control range:25ºC~37ºC
Skin Temperature control range:34ºC~37ºC
Sensor precision: ≤0.3ºC
Changing of temperature: ≤0.5ºC
Uniformity of temperature: ≤0.8ºC
Temperature rising time: ≤30min
Noise level: ≤55dB(A)
Inclination of bassinet: 0°~5°
Alarming:
audible and visible alarm indication, and it can indicate the real temperature and cut off the power supply of heater (except for the low deviation alarm), meanwhile, the skin temperature indication window can indicate the alarm code. For the alarm 1-6, the relevant light is on, and for the alarm 7, all alarming lights are on; all alarm sound can be cancelled except for the power failure alarm and the system alarm.
Over-temp alarm: when the air temperature control ≤38ºC, skin temperature control≤40ºC;
Upper Deviation alarm: air temperature control +3.0ºC, skin temperature control +1.0ºC;
Lower Deviation alarm: air temperature control -3.0ºC, skin temperature control -1.0ºC;
Sensor alarm: when the sensor is disconnected, open circuit, short circuit or put on the wrong place;
Fan alarm: when the fan is blocked, speed ≤1000rpm or stop working;
Power failure alarm: when there is no power or the power cord is disconnected.
System alarm: when there is E2ROM malfunction inside of temperature controller.