LCD Multi-Parameter Patient Monitor (PM-300A)

  • Model NO.:PM-300A

  • Type:Pulse Oximeter

  • Group:All People

  • Gross Weight:4.5kg

  • Power Voltage:AC 100 V-240V 50-60Hz

  • Temperature Operation:0 ~ 40 °c

  • Safety Level:Class , Type CF

  • Specification:CE, ISO

  • HS Code:90181930

  • Classification:Imaging Diagnostic Equipment

  • Certification:CE

  • Net Weight:3.5kg

  • Package Size:258*210*175mm

  • Power Iutput:≤70va

  • Fuse:1.6A

  • Trademark:everich

  • Origin:China

Product Description

Features:  
      7 inch high resolution (800x600) colorful TFT LCD
           Simultaneous display of 4 channel waveform on one screen
           Dual NIBP with over pressure protection
           Analysis of real time ST segment and arrhythmia
           Standard configuration: ECG, NIBP, RESP, SPO2, 1-TEMP, PR, drug dose calculation, OXYCRG
           Optional configuration: Built in thermal printer, 2-TEMP
           Standard RJ-45 socket to support networking capacity of maximum 66 beds and compatibility with HIS
           Digital SPO2 enables effective examination in case of active patient and low infusion as well
                interference caused by electrosurgical unit
           Compact and portable design allows uninterrupted operation with its built in rechargeable battery
Specifications:

Size and Weight



                     Name of the monitor


                     Multi-parameter Monitor


                     Size


                     258mm×210mm×175mm


                     Weight


                     4.5Kg


             System output



                     Display type


                     12.1″LCD color display screen, with a resolution of 800×600


                     Interface with central unit


                     Network port


             Power supply



                     Power Voltage


                     AC 100V~240V 50/60Hz


                     Power Input


                     ≤70VA


                     Fuse


                     1.6A


                     Safety Level


                     Class I , Type CF


             Battery



                     Type


                     Rechargeable Lead acid cell, 12V/3.2AH


                     Charge time


                     ≤6 hours


                     Working hours when charged fully


                     ≥30min when using normally


             Environment



                     Ambient Temperature


                     Working 0~+40ºC

                     Transportation and storage -20~+50ºC


                     Relative humidity


                     Working ≤85%

                     Transportation and storage ≤93%


                     Atmospheric pressure


                     Working 860~1060 hPa

                     Transportation and storage 500~1060 hPa


             ECG



                     Lead Mode


                     5-leads ECG input

                     3-leads ECG input


                     Lead selection


                     I, II, III, aVR, aVL, aVF, V-

                     I, II, III

                     I, II, III, aVR, aVL, aVF, V1~V6


                     Gain


                     2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV


                     Input impedance


                     ≥5.0 Mohm(d.c. ~ 60 Hz)


                     CMRR


                     DIAG ≥85dB

                     MON ≥105dB

                     OPS ≥105dB


                     Frequency response


                     DIAG 0.05~75Hz

                     MON 0.5~25Hz

                     OPS 1~15Hz


                     Electrode offset potential


                     ±300 mV d.c. → ±600mV d.c.


                     Leakage Current


                     <10 uA


                     ECG signal range


                     ±6.0 mV


                     Baseline recovery


                     <5 s After Defi. (MON or OPS mode)


                     Insulation


                     Breakdown Voltage 4000VAC 50Hz/60Hz


                     Indication of electrode separation


                     Available


                     Sweep speed


                     12.5mm/s, 25mm/s, 50mm/s


             HR Measuring and Alarm



                     Range


                     10~300 bpm


                     Refreshing time


                     Per pulse


                     Resolution


                     1 bpm


                     Accuracy


                     ±1% or ±1 bpm, whichever is greater


                     Alarm range


                     10~300 bpm, continuously adjustable between high limit and low limit


                     Alarm indication


                     Sound and light alarming


             NIBP



                     Way of measurement


                     Automatic oscillometry


                     Range of measurement


                     Adult


                     SYS


                     30~275 mmHg


                     DIA


                     10~255 mmHg


                     MEAN


                     20~255 mmHg


                     Child


                     SYS


                     30~235 mmHg


                     DIA


                     10~160 mmHg


                     MEAN


                     20~170 mmHg


                     Neonate


                     SYS


                     30~145 mmHg


                     DIA


                     10~120 mmHg


                     MEAN


                     20~140 mmHg


                     Cuff pressure range


                     0~280 mmHg


                     Resolution


                     1 mmHg


                     Pressure Accuracy

                     Static

                     Clinical


                     ±2% or ±3 mmHg, whichever is greater

                     ±5 mmHg average error

                     8 mmHg standard deviation


                     Unit


                     mmHg, kPa


                     Pulse rate measuring range


                     40 ~ 240 bpm


                     Inflation time for cuff


                     Less than 40 sec. (standard adult cuff)


                     Total cycle time


                     20 to 45 seconds typical (dependent on heart rate and motion artifact)


                     Intervals for cycle measurement time


                     1,2,3,4,5,10,15,20,30,45,60,90 minutes

                     2,4,8 hours


                     Overpressure Protection

                     Adult

                     Child

                     Neonatal


                     Hardware and software double protections

                     315±10 mmHg

                     255±10 mmHg

                     170±10 mmHg


                     Range of alarm


                     SYS


                     0~300 mmHg, continuously adjustable between high limit and low limit


                     DIA


                     0~300 mmHg, continuously adjustable between high limit and low limit


                     MEAN


                     0~300 mmHg, continuously adjustable between high limit and low limit


                     Alarm indication


                     Sound and light alarming


                     Ways of measurement


                     Single, Cycle and Quick (Q.sys)


             SpO2



                     Range of measurement


                     0~100%


                     Resolution


                     1%


                     Precision


                     At 70~100%, ± 2%

                     At 35~69%, ± 3%

                     At 0~34%, unspecified


                     Range of alarm


                     0~100%, continuously adjustable between high limit and low limit


                     Alarm indication


                     Sound and light alarming


                     Type of sensor


                     Dual bandwidth light emitting diode


                     PR:

                     Measuring and alarm Range

                     Resolution

                     Accuracy


                     25~250 bpm

                     1 bpm

                     ±2% or ±1 bpm, whichever is greater



TEMP



                     Range of measurement


                     0.0~50.0ºC


                     Resolution of display


                     0.1ºC


                     Unit


                     Celsius (ºC), Fahrenheit (ºF)


                     Precision


                     ±0.1ºC


                     Connecting cable


                     Compatible with YSI-4400


                     Range of alarm


                     0.0~50.0ºC, continuously adjustable between high limit and low limit


                     Alarm indication


                     Sound and light alarming


             RESP



                     Method


                     Impedance between RA-LL (R-F)


                     Measuring impedance range


                     0.5 ~3 Ohms


                     Base line impedance range


                     500~4000 Ohms (50~120 kHz exciting frequency)


                     Range of measurement


                     0~120 bpm


                     Precision


                     ±2 bpm


                     Gain


                     x1,x2,x4


                     Range of alarm


                     0~120 bpm, continuously adjustable between high limit and low limit


                     Sweep speed


                     6.25mm/s, 12.5mm/s, 25mm/s


                     Asphyxia


                     1~60s, off


                     Alarm indication


                     Sound and light indication


             Recorder



                     Speed of paper feeding


                     12.5mm/s, 25mm/s, 50 mm/s (Only valid whit B Type recorder)


                     Recording medium


                     50mm thermal paper


                     Recording track


                     Single track or tri-track