LCD Multi-Parameter Patient Monitor (PM-200A)

  • Model NO.:PM-200A

  • Type:Pulse Oximeter

  • Group:All People

  • Gross Weight:4.5kg

  • Power Voltage:AC 100 V-240V 50-60Hz

  • Temperature Operation:0 ~ 40 °c

  • Resolution:1bpm

  • Specification:CE, ISO

  • HS Code:90181930

  • Classification:Imaging Diagnostic Equipment

  • Certification:CE

  • Net Weight:3.5kg

  • Package Size:318*264*152mm

  • Power Iutput:≤70va

  • Measurement Range:10-300bpm

  • Trademark:everich

  • Origin:China

Product Description

Features:  
    12.1 inch high brightness colorful TFT LCD
            Simultaneous display of 7 channel waveform by one screen
            Dual NIBP with over pressure protection
            Analysis of real time ST segment and arrhythmia
            Standard configuration: ECG, NIBP, RESP, SPO2, 1-TEMP, PR, drug dose calculation, OXYCRG
            Optional configuration: 12 lead ECG, 2-TEMP, IBP(single/dual channel), ETCO2(side stream/main stgream),
                built in thermal printer and CO
            Standard RJ-45 socket to support networking capacity of maximum 66 beds and compatibility with HIS
            Digital SPO2 enables effective examination in case of active patient and low infusion as well
                interference caused by electrosurgical unit
            Compact and portable design allows uninterrupted operation with its built in rechargeable battery
Specifications:

Size and Weight
 

Name of the monitor

Multi-parameter Monitor

Size

318mm×264mm×152mm

Weight

4.5Kg

System output
 

Display type

12.1″LCD color display screen, with a resolution of 800×600

Interface with central unit

Network port

Power supply
 

Power Voltage

AC 100V~240V 50/60Hz

Power Input

≤70VA

Fuse

1.6A

Safety Level

Class I , Type CF

Battery
 

Type

Rechargeable Lead acid cell, 12V/3.2AH

Charge time

≤6 hours

Working hours when charged fully

≥30min when using normally

Environment
 

Ambient Temperature

Working 0~+40ºC

                     Transportation and storage -20~+50ºC

Relative humidity

Working ≤85%

                     Transportation and storage ≤93%

Atmospheric pressure

Working 860~1060 hPa

                     Transportation and storage 500~1060 hPa

ECG
 

Lead Mode

  1. 5-leads ECG input

  2. 3-leads ECG input

  3. 10-leads ECG input

Lead selection

  1. I, II, III, aVR, aVL, aVF, V-

  2. I, II, III

  3. I, II, III, aVR, aVL, aVF, V1~V6

Gain

2.5mm/mV, 5mm/mV, 10mm/mV, 20mm/mV, 40mm/mV

Input impedance

≥5.0 Mohm (d.c. ~ 60 Hz)

CMRR

DIAG ≥85dB

                     MON ≥105dB

                     OPS ≥105dB

Frequency response

DIAG 0.05~75Hz

                     MON 0.5~25Hz

                     OPS 1~15Hz

Electrode offset potential

±300 mV d.c. → ±600mV d.c.

Leakage Current

<10 uA

ECG signal range

±6.0 mV

Baseline recovery

<5 s After Defi. (MON or OPS mode)

Insulation

Breakdown Voltage 4000VAC 50Hz/60Hz

Indication of electrode separation

Available

Sweep speed

12.5mm/s, 25mm/s, 50mm/s

HR Measuring and Alarm
 

Range

10~300 bpm

Refreshing time

Per pulse

Resolution

1 bpm

Accuracy

±1% or ±1 bpm, whichever is greater

Alarm range

10~300 bpm, continuously adjustable between high limit and low limit

Alarm indication

Sound and light alarming

NIBP
 

Way of measurement

Automatic oscillometry

Range of measurement

Adult

SYS

30~275 mmHg

DIA

10~255 mmHg

MEAN

20~255 mmHg

Child

SYS

30~235 mmHg

DIA

10~160 mmHg

MEAN

20~170 mmHg

Neonate

SYS

30~145 mmHg

DIA

10~120 mmHg

MEAN

20~140 mmHg

Cuff pressure range

0~280 mmHg

Resolution

1 mmHg

Pressure Accuracy

                     Static

                     Clinical

±2% or ±3 mmHg, whichever is greater

                     ±5 mmHg average error

                     8 mmHg standard deviation

Unit

mmHg, kPa

Pulse rate measuring range

40 ~ 240 bpm

Inflation time for cuff

Less than 40 sec. (standard adult cuff)

Total cycle time

20 to 45 seconds typical (dependent on heart rate and motion artifact)

Intervals for cycle measurement time

1,2,3,4,5,10,15,20,30,45,60,90 minutes

                     2,4,8 hours

Overpressure Protection

                     Adult

                     Child

                     Neonatal

Hardware and software double protections

                     315±10 mmHg

                     255±10 mmHg

                     170±10 mmHg

Range of alarm

SYS

0~300 mmHg, continuously adjustable between high limit and low limit

DIA

0~300 mmHg, continuously adjustable between high limit and low limit

MEAN

0~300 mmHg, continuously adjustable between high limit and low limit

Alarm indication

Sound and light alarming

Ways of measurement

Single, Cycle and Quick (Q.sys)

SpO2
 

Range of measurement

0~100%

Resolution

1%

Precision

At 70~100%, ± 2%

                     At 35~69%, ± 3%

                     At 0~34%, unspecified

Range of alarm

0~100%, continuously adjustable between high limit and low limit

Alarm indication

Sound and light alarming

Type of sensor

Dual bandwidth light emitting diode

PR:

                     Measuring and alarm Range

                     Resolution

                     Accuracy

25~250 bpm

                     1 bpm

                     ±2% or ±1 bpm, whichever is greater

IBP
 

Range of measurement

-45 ~ +300 mmHg

Resolution

1 mmHg

Precision

± 4mmHg or 4% of the reading (select the bigger one of the two)

Sensitivity of transducer

5uV/V/mmHg, 2%

Impedance of transducer

300~3000 Ohm

Bandwidth

d.c. ~ 15Hz

Transducer sites

Arterial Pressure (ART)

                     Pulmonary Artery Pressure (PA)

                     Left Atrium Pressure (LAP)

                     Right Atrium Pressure (RAP)

                     Central Venous Pressure (CVP)

                     Intracranial Pressure (ICP)

Selection of measurement range

ART

0~50mmHg

                     50~150 mmHg

                     100~240 mmHg

                     0~300 mmHg

                     AUTO

PA

                     CVP

                     LAP

                     RAP

                     ICP

0~20 mmHg

                     0~30 mmHg

                     0~50 mmHg

                     0~80 mmHg

                     AUTO

(Among them, the AUTO switches automatically at an interval of 10 mmHg so as to ensure the waveform is at the state most suitable for observation)

Range of alarm

SYS

-10~300 mmHg, continuously adjustable between high limit and low limit

DIA

-10~300 mmHg, continuously adjustable between high limit and low limit

MEAN

-10~300 mmHg, continuously adjustable between high limit and low limit

Alarm indication

Sound and light alarm

Unit

mmHg, kPa

Insulation resistance

>5000MΩ/1000V

Breakdown voltage

>4000V AC 50Hz/60Hz

TEMP
 

Range of measurement

0.0~50.0ºC

Resolution of display

0.1ºC

Unit

Celsius (ºC), Fahrenheit (ºF)

Precision

±0.1ºC

Connecting cable

Compatible with YSI-4400

Range of alarm

0.0~50.0ºC, continuously adjustable between high limit and low limit

Alarm indication

Sound and light alarming

RESP
 

Method

Impedance between RA-LL (R-F)

Measuring impedance range

0.5 ~3 Ohms

Base line impedance range

500~4000 Ohms (50~120 kHz exciting frequency)

Range of measurement

0~120 bpm

Precision

±2 bpm

Gain

x1,x2,x4

Range of alarm

0~120 bpm, continuously adjustable between high limit and low limit

Sweep speed

6.25mm/s, 12.5mm/s, 25mm/s

Asphyxia

1~60s, off

Alarm indication

Sound and light indication

Recorder
 

Speed of paper feeding

12.5mm/s, 25mm/s, 50 mm/s (Only valid whit B Type recorder)

Recording medium

50mm thermal paper

Recording track