Medical Bubble CPAP Ventilation System (NLF-200D)

  • Model NO.:NLF-200D

  • Continuous Flow Adjusting Range:1.5 L/Min ~ 15 L/Min

  • CPAP Monitoring Range:0 CMH2o ~ 60 CMH2o

  • Rated Working Pressure:350 Kpa

  • Gas Source:O2 / Air (for Medical Use)

  • Specification:CE, ISO

  • HS Code:90189090

  • CPAP Adjusting Range:0 CMH2o ~ 10 CMH2o

  • Oxygen Concentration:21%-100%

  • Continuous Flow Monitoring Range:1.5 L/Min ~ 15 L/Min

  • Output Pressure of Regulator:300 Kpa~400 Kpa

  • Trademark:everich

  • Origin:China

Product Description

Application
CPAP system is used for newborn baby with stronger spontaneous breath and slightly respiratory failure, and applies to pediatric clinics for NCPAP oxygen uptake.

Performance and features
1. CPAP ventilation modes keeps children spontaneous breath and reduces children breath work
2. Continues to provide CPAP mode of constant flow and limited pressure for newborn baby
3. Delivery air/O2 mixture with properly adjusted O2 concentration and continuous positive pressure to premature baby, newborn baby and child which will be highly improve the safety and efficiency of oxygen therapy
4. Equipped with build-in air blender, ensuring O2 concentration adjustment to be stable.
5. Reliable CPAP control valve and pressure monitoring system improve CPAP adjustment precision
6. Optional parts: humidifier with heating devices
7. High cost-effective and reliable quantity
8. Easy operation, compact appearance and design let user could operate more easily.
 
Technical parameters:

1)Parameter adjustment
CPAP adjusting range                                  0 cmH2O ~ 10 cmH2O.
Continuous flow adjusting range                    1.5 L/min ~ 15 L/min
Oxygen Concentration                    21%-100%
 

2)Parameter monitoring range
CPAP/airway pressure
--Monitoring range                                 0 cmH2O ~ 60 cmH2O
--Accuracy                                          ±(2 % full scale + 4 % actual value)
Continuous flow
--Monitoring range                                    1.5 L/min ~ 15 L/min
--Accuracy                                              ±4 %
 

3) Gas supply
Gas source                                            O2(For medical use)
                                 Air(For medical use)
Rated working pressure                   350 kPa
Output pressure of regulator            350 kPa±50 kPa (300 kPa~400 kPa)
 
 
 
Work environment:
Ambient temperature                 5 ºC~ 40 ºC
Relative humidity                          ≤80 %
Atmospheric pressure                 960 hPa~1040 hPa
Storage and transportation environment
Ambient temperature                 -10 ºC~+40 ºC
Relative humidity                          ≤90 %
Atmospheric pressure                 860 hPa~1060 hPa



CPAP adjusting rangeContinuous flow adjusting rangeOxygen ConcentrationCPAP Monitoring rangeContinuous flow Monitoring rangeRated working pressureOutput pressure of regulatorGas source
0 cmH2O ~ 10 cmH2O1.5 L/min ~ 15 L/min21%-100%0 cmH2O ~ 60 cmH2O1.5 L/min ~ 15 L/min350 kPa300 kPa~400 kPaO2 / Air (For medical use)