Nlf-200c CPAP Ventilation System Machine

  • Model NO.:NLF-200C

  • Power Supply (with Air Compressor):~220 V±22 V, 50 Hz

  • Air Source:O2 / Air (for Medical Use)

  • Input Power(Without Air Compressor):250va

  • Specification:CE, ISO

  • HS Code:90189090

  • Power Supply (Without Air Compressor):100-240V,50/60Hz

  • Air Source Pressure:280kpa ~ 600kpa

  • Input Power(with Air Compressor):900va

  • Trademark:everich

  • Origin:China

Product Description

Application
CPAP system is used for newborn baby with stronger spontaneous breath and slightly respiratory failure, and applies to pediatric clinics for NCPAP oxygen uptake
 
Performance and features
1.      CPAP ventilation mode keeps children spontaneous breath and reduces children breathing work, continues to provide CPAP mode of constant flow and limited pressure for newborn baby;
2.      Delivery air/O2 mixture with properly adjusted O2 concentration and continuous positive pressure to premature baby, newborn baby and child, which will highly improve the safety and efficacy of oxygen therapy;
3.      Equipped with built-in oxygen/air blender, ensuring O2 concentration adjustment to be stable ;
4.      High quality oxygen sensor can monitor real-time O2 concentration , alarm over limitation, ensure safety of breathing O2;
5.      Reliable CPAP control valve and advanced pressure monitoring system improve CPAP adjustment precision.
6.      3'' display
 
Working condition
l  Air source                                        O2,AIR
l  Air source pressure               280kPa ~ 600kPa
l  Power supply                                  ~220 V±22 V, 50 HZ with air compressor
                                       100-240V,50/60HZ(without air compressor)
l  Input power                                        900VA(with air compressor)
                                    250VA (without air compressor)
 
Ventilation Mode
l  CPAP

Main technical parameters
Adjusting Parameter
l  Inhaled oxygen concentration    21%~100% 
l  CPAP                         0cmH2O~15 cmH2O  
l  Continuous flow               2L/min~20L/min
Monitoring Parameter
l  Inhaled oxygen concentration         15 % ~ 100 %
l  CPAP                         0 cmH2O ~ 60 cmH2O
l  Continuous flow                2 L/min ~ 20 L/min

Alarm and protection
l  The AC power cut-off alarm
l  Low air (oxygen) input pressure alarm
l  Air / oxygen input unbalanced pressure alarm
l  High oxygen concentration alarm
l  low oxygen concentration alarm
l  Pressure limit

Power supply Input powerAir source pressureAir source
with air compressorwithout air compressorwith air compressorwithout air compressor
100-240V,50/60HZ~220 V±22 V, 50 HZ900VA250VA280kPa ~ 600kPaO2 / Air (For medical use)