Application
CPAP system is used for newborn baby with stronger spontaneous breath and slightly respiratory failure, and applies to pediatric clinics for NCPAP oxygen uptake
Performance and features
1. CPAP ventilation mode keeps children spontaneous breath and reduces children breathing work, continues to provide CPAP mode of constant flow and limited pressure for newborn baby;
2. Delivery air/O2 mixture with properly adjusted O2 concentration and continuous positive pressure to premature baby, newborn baby and child, which will highly improve the safety and efficacy of oxygen therapy;
3. Equipped with built-in oxygen/air blender, ensuring O2 concentration adjustment to be stable ;
4. High quality oxygen sensor can monitor real-time O2 concentration , alarm over limitation, ensure safety of breathing O2;
5. Reliable CPAP control valve and advanced pressure monitoring system improve CPAP adjustment precision.
6. 3'' display
Working condition
l Air source O2,AIR
l Air source pressure 280kPa ~ 600kPa
l Power supply ~220 V±22 V, 50 HZ with air compressor
100-240V,50/60HZ(without air compressor)
l Input power 900VA(with air compressor)
250VA (without air compressor)
Ventilation Mode
l CPAP
Main technical parameters
Adjusting Parameter
l Inhaled oxygen concentration 21%~100%
l CPAP 0cmH2O~15 cmH2O
l Continuous flow 2L/min~20L/min
Monitoring Parameter
l Inhaled oxygen concentration 15 % ~ 100 %
l CPAP 0 cmH2O ~ 60 cmH2O
l Continuous flow 2 L/min ~ 20 L/min
Alarm and protection
l The AC power cut-off alarm
l Low air (oxygen) input pressure alarm
l Air / oxygen input unbalanced pressure alarm
l High oxygen concentration alarm
l low oxygen concentration alarm
l Pressure limit
Power supply | Input power | Air source pressure | Air source |
with air compressor | without air compressor | with air compressor | without air compressor |
100-240V,50/60HZ | ~220 V±22 V, 50 HZ | 900VA | 250VA | 280kPa ~ 600kPa | O2 / Air (For medical use) |