Brief introduction:
This unit is used for transporting the newborn babies between sickrooms or hospitals; AC Power and DC Power can be used alternatively and the device can be connected with Power 12V or 24V. Air-temp or skin-temp servo-controlled by micro-computer; Double walls hood could prevent heat losing from hood and prevent the influence to the incubator from exterior heat resource; Quite operation windows can reduce the disturbance to the baby during nursing with oxygen cylinder and oxygen supply system; ; Touch switch panel with LED digital display, set-temp, skin-temp, real-temp, storage battery voltage are displayed separately; Indicated electricity-offer mode; Night light, luminescence design; Seven self-checking alarms with audible and visible; Heating power rate indicates 0-100%; Shockproof and adjustable stand with cylinder and three steps height.
Brief introduction:
Two temperature control modes: Air mode and Baby mode controlled by micro-computer; Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance;
Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately;
Failure alarm indication;
Can be equipped with >37ºC temperature set function;
Double wall Acrylic glass hood with side door so that the bassinet can be pulled out from it;
Natural air flow humidity;
The height of whole unit can be adjusted;
Oxygen supply system;
Observation lamp
Disposable skin temperature sensor is optional;
Ambulance trolley is optional
Standard configuration:
Main body (including the hood, controller, bassinet, internal battery, and observation lamp), oxygen cylinders, oxygen supply system, skin temperature sensor, I.V. Pole, mattress, adjustable stand)
Optional configuration: >37ºC override temperature mode
Specification
AC Power supply | AC220V-230V/50Hz or AC110-120V/50-60Hz |
DC Power supply | DC12V/10A or DC24V/6A |
Input power | 400VA |
Mode of control | Air mode and baby mode controlled by micro-computer |
Air temperature control range | 25ºC-37ºC(Override mode 37-38ºC) |
Baby temperature control range | 34ºC-37ºC(Override mode 37-37.5ºC) |
Skin temperature sensor accurac | ±0.3ºC |
Temperature variability | ≤1.0ºC |
Temperature uniformity | ≤1.5ºC(mattress is on the horizontal position) |
Noise inside hood | ≤52 dB(A) |
Operating range | 10 ºC~30ºC |
Ambient air movement rate | <1.0m/s |
AC Power supply: AC220V-230V/50Hz or AC110-120V/50-60Hz
DC Power supply: DC12V/10A or DC24V/6A
Input power: 400VA
Mode of control: Air mode and baby mode controlled by micro-computer
Air temperature control range: 25ºC-37ºC(Override mode 37-38ºC)
Baby temperature control range: 34ºC-37ºC(Override mode 37-37.5ºC)
Skin temperature sensor accuracy: ±0.3ºC
Temperature variability: ≤1.0ºC
Temperature uniformity: ≤1.5ºC(mattress is on the horizontal position)
Noise inside hood: ≤52 dB(A)
Failure alarm: over-temperature alarm, temperature deviation alarm, sensor alarm, Motor Fan Failed Alarm, Power failure Alarm and so on.
Operating at temperature equilibrium:90min(Set Temp 36ºC-Ambient 15ºC)
Environmental temperature
Operating range: 10 ºC~30ºC
(Set temperature should be higher 3ºC than the environment temperature)
Ambient air movement rate: <1.0m/s
Packing
Package: each unit is packed into two cartons
Dimension (Main body): 109cm x 65.5cm x 76cm
(Cabinet): 138cm x 61cm x 40cm
Gross Weight (Main body): 80Kg
(Cabinet): 53Kg